Pre-Bypass Checklist
Patient
-
Chart reviewed
-
Procedure reviewed
-
Blood and/or Components available
Sterility
-
Disposable equipment verified for sterility and integrity
-
Heat Exchangers Water Leak tested
Pump
-
Speed controls operational
-
Roller heads engage tubing properly, smooth & quiet rotations
-
Biohead Engaged
-
Occlusions set and verified
-
Flow meter in correct direction and calibrated
-
Flow rate indicator correct for patient and/or tubing size
-
All Holders secured to posts and cross beams
Electrical
-
Power cord(s) securely connected
-
Cord in good condition ( no visible damage)
-
UPS/Battery Back-up Charged and Operational
Gas Supply
-
Gas line checked for kinks and patency
-
Flow meter/ blender functional
-
Gas exhaust unobstructed
-
Vaporizer in working order and full
Circuit and Tubing
-
Connections secure
-
Tubing direction traced and correct
-
No kinks noted
-
One-way valve in correct direction
-
Circuit De-aired and Pressure/Flow Checked
-
Arterial filter Debubbled
-
Recirculation Line Checked
-
Cardiotomy reservoir vented
-
Oxygenator Failure Checklist
Cardioplegia System
-
Solutions check for expiration and added substrates
-
System De-aired and Pressure Checked
-
Connections secure and leak free
Monitoring
-
Temperature Probes Connected
-
Pressure Monitors Alarms On and Calibrated
-
(Optional) Oxygen analyzer calibrated
-
Saturation Monitors or Blood Analyzers calibrated
Temperature Control
-
Water source connected and functional
-
Water baths full
Supplies
-
Tubing clamps available
-
Drugs available
-
Solutions available
-
Sampling syringes/ laboratory tubes available
Anticoagulation
-
Anticoagulation Controls within calculated parameters
-
Warming Block at correct temperature
Emergency equipment
-
Hand cranks available
-
Flashlight or Emergency lighting available
-
Duplicate circuit components available
-
Alarms operational and engaged